Twenty milliseconds is all it takes, and researchers say every millisecond of resolution counts. 

Keratoconus may start changing how the cornea behaves before those changes become visible on conventional imaging. A prototype optical coherence tomography (OCT) system that measures the cornea’s response to an air puff at nine locations simultaneously could catch it sooner, according to a proof-of-concept study published in Biomedical Optics Express.

Measuring nine points at once 

The prototype OCT system tracks the cornea’s response to an air puff at one central and eight peripheral locations at the same time. Conventional systems typically assess only a single point or a median value. Instead, this approach maps localized biomechanical changes across the cornea during the same deformation event.*

“The corneal deformation we aim to capture…lasts only about 20 milliseconds, and the measurement cannot be repeated,” said the study’s first author, Dr. Karol Karnowski, in a news release. “This makes the simultaneous acquisition of the corneal response at multiple locations the key feature of our method. Achieving this requires very high temporal resolution.”

The team also introduced “asymmetry vector,” a metric that summarizes both the magnitude and direction of biomechanical imbalance.* 

READ MORE: Keratoconus Management Gets New Options at ASCRS 2026

Finding the weakest spots 

The asymmetry vector consistently pointed toward areas of localized corneal weakening. In eyes with keratoconus, it aligned with the region of thinning, steepening and abnormal posterior elevation seen on corneal tomography, suggesting the measurements captured the same disease-affected areas.*

The system also showed good repeatability in healthy eyes. When investigators simulated slower acquisition speeds, measurement accuracy dropped substantially, underscoring the importance of capturing the cornea’s rapid mechanical response in real time.*

The study also turned up an unexpected finding: a previously unreported “dual-indentation” response, in which the cornea underwent two distinct deformation events after a modified air-puff stimulus. The biological mechanism behind the phenomenon remains unclear and will require further investigation.*

WATCH: EpiSmart Cross-Linking Earns FDA Fast Track for EpiOn Keratoconus Care

Limitations

With only 13 healthy volunteers and three keratoconus patients, the study was not designed to establish clinical performance. The authors said larger studies are needed to validate the technology, and the current prototype still relies on manual image analysis that isn’t yet practical for routine clinical use.*

READ MORE: Glaukos Launches Epioxa: First FDA-Approved Incision-Free Treatment for Keratoconus

The takeaway 

The study expands what OCT may be able to measure, adding localized biomechanical behavior to conventional structural assessment. 

The cornea, it turns out, has more to say than its shape alone lets on—whether that translates into earlier diagnosis or better clinical decision-making remains to be established in larger studies.

*Karnowski K, Milkiewicz J, Cetinkaya O, et al. Single-shot, depth-encoded multiplexed OCT for multi-spot tracking of induced transient corneal dynamics. Biomedical Optics Express. 2026;17(5):2715–2733. 